QUICK SET-UP

Model No. 900-340
Electronic Superficial Rockwell tester

(HRM-45DT)
  1. Loosen 4-nuts on bottom of crate. Lift the crate off of the platform being careful not to scratch side of tester.
  2. Remove the tool kit from the platform.
  3. Remove plastic dust cover
  4. Remove string holding handles and top cover
  5. Remove 2-phillips head screws from front of face above the dial; remove 1-phillips head screw from back of top cover.   Lift top cover straight up.
  6. Remove rubber band holding indicator lever and discard
  7. Remove back cover by removing 2-phillip head screws from top/bottom
  8. Carefully remove strings that are holding the weight bar to the loading arm
  9. Remove the weight bar from the loading arm by lifting the hook of the bar from the loop on the arm.
  10. Remove the 2-nuts at the bottom of the bar.
  11. Open test kit and remove the 3 weights. Clean any oil/grease from weights
  12. Take the round weight marked “1” and slide up the bottom of the weight bar and proceed to put on both nuts.
  13. Carefully place weight bar back in loop of loading arm with the hook facing the front of the machine.
  14. Place weight “2” and place on the middle section of the weight bar keeping the flat part to the rear of the machine. Place weight “3” on the top part of the weight bar/flat to rear.
  15. Install rear cover with 2-screws
  16. Install top cover (place straight down) with 2-screws on the front face and 1-screw on the rear part of the top cover.
  17. Lower rotary handle and remove black plastic shaft protector. Save this piece for safety.
  18. Choose applicable work table and clean off grease. Install on machine.
  19. Remove diamond penetrator from case and install in penetrator shaft making sure the flat of the penetrator is facing the set screw. Be sure penetrator contacts the bottom of shaft flush. Do not overtighten!
  20. Set weight selector on correct weight.
  21. Turn on power. The light will shine on the test surface.
  22. Place the test block on the anvil and rotate handle until the block touches the penetrator. At that time, continue turning the handle 3 times around until the needle reaches the “zero” at TDC.
  23. Push the white button on the right side of the base. The tester will automatically put the weight on. When the weight is on, the light will go out for a few seconds. At that point, the light will go back on and the machine will automatically release the weight. When the weight release procedure is done, you can take your reading at that time.
  24. Rotate the handle to lower the test block from the penetrator.
 
 
HARDNESS TESTER RESOURCES
Hardness Tester Accessories Portable Hardness Testers
Hardness Tester Block Certification Minimum Thickness Chart for Rockwell Hardness Testing
Rebuilt Inventory with Full Warranty Rockwell Hardness Scale Indentor Load Chart
Hardness Tester Conversion Chart Hardness Setup:
  • 900-331
  • 900-340
  • 900-360
  • 900-370
  • Principle of Rockwell Hardness Testing NIST Links
    Roundness Correction Chart for Hardness Testing ASTM Links
    HARDNESS TESTER LINKS
    TRADE SHOWS
    • Visit our booth #6824 @ Quality Expo September 25-27, Rosemont, Illinois